Test Code LAB123367 Titanium, Serum
Additional Codes
Mayo Test ID |
---|
TIS |
Reporting Name
Titanium, SUseful For
Monitoring metallic prosthetic implant wear
Performing Laboratory
Mayo Clinic Laboratories in RochesterSpecimen Type
SerumSpecimen Required
Patient Preparation: High concentrations of gadolinium and iodine are known to interfere with most metal tests. If either gadolinium- or iodine-containing contrast media has been administered, a specimen should not be collected for at least 96 hours.
Supplies:
-Metal Free Specimen Vial (T173)
-Metal Free B-D Tube (No Additive), 6 mL (T184)
Collection Container/Tube: Plain, royal blue-top Vacutainer plastic trace element blood collection tube
Submission Container/Tube: 7-mL Metal-free, screw-capped, polypropylene vial
Specimen Volume: 1.2 mL
Collection Instructions:
1. Allow specimen to clot for 30 minutes; then centrifuge the specimen to separate serum from the cellular fraction.
2. Remove the stopper. Carefully pour specimen into a metal-free, polypropylene vial, avoiding transfer of the cellular components of blood. Do not insert a pipet into the serum to accomplish transfer, and do not ream the specimen with a wooden stick to assist with serum transfer.
3. See Metals Analysis Specimen Collection and Transport for complete instructions.
Specimen Minimum Volume
0.4 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Serum | Refrigerated (preferred) | 28 days | METAL FREE |
Frozen | 28 days | METAL FREE | |
Ambient | 7 days | METAL FREE |
Special Instructions
Reference Values
<2 ng/mL
Day(s) Performed
Wednesday
Test Classification
This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.CPT Code Information
83018
LOINC Code Information
Test ID | Test Order Name | Order LOINC Value |
---|---|---|
TIS | Titanium, S | 8244-6 |
Result ID | Test Result Name | Result LOINC Value |
---|---|---|
89367 | Titanium, S | 8244-6 |
Report Available
2 to 8 daysReject Due To
Gross hemolysis | OK |
Gross lipemia | OK |
Gross icterus | OK |
Method Name
Triple-Quadrupole Inductively Coupled Plasma Mass Spectrometry (ICP-MS)